Can we Assist you?
Please feel free to contact us.
Sulfur and nitrogen analyzer trace SN cube
trace SN cube
Sulfur and nitrogen analyzer
ll_to Content

The very best in ppb analysis

Meet the most sensitive combustion analyzer for sulfur and nitrogen in diesel, gasoline, and related samples: the trace SN cube. This sulfur and nitrogen analyzer with its catalyst-supported combustion technology ensures complete decomposition of all kinds of samples without soot formation and delivers fast analysis results in just a few minutes. By incorporating two complete combustion and detection channels for sulfur and nitrogen, respectively, into one analyzer, the trace SN cube works as two analyzers in one box. This concept allows the specific optimization of the determination for each element. This ensures sulfur and nitrogen analysis with highest precision and accuracy. The analyzer combines highest analytical performance with speed, flexibility, and long-term stability in automated operation.

Highlights of the trace SN cube

The most sensitive sulfur and nitrogen determination

with detection limits of 6 ppb for sulfur and 15 ppb for nitrogen

Soot-free combustion

for reliable results and maintenance-friendly instrument operation

Effortless Operation

simplifies daily routine with a low-maintenance design

Highest sample flexibility

Optionally available modules allow the analysis of solids, gases, liquefied petroleum gas (LPG), or chlorine

Application Guide Organic Elemental Analysis

If you can combust it, you can measure it.

Whether you're working with pharmaceuticals, chemicals, soils, biomass, fuels, or advanced materials, elemental analysis provides the critical data needed to understand your samples and ensure reliable results. Our comprehensive Application Guide brings together application expertise, analytical know-how, and real-world case studies from multiple industries. 

Cover of the organic elemental analysis application guide, incl. a man in the lab, weighing samples
DOWNLOAD YOUR COPY

Fill in the form to receive your download link per e-mail.

Your contractual consideration for the free provision of the download is the subscription to our personalized newsletter. By clicking on the “download now” button, you therefore declare your acceptance of the receipt of personalized newsletters by e-mail by Elementar India Pvt Limited and its group companies as well as the evaluation of your user behavior in this regard and - if available - the merging of this data with your data in our customer database.

In order to receive newsletters from our group companies it is necessary to transfer your above-mentioned personal data to these companies. The data transfer is contractually required.

You are aware that the subscription to our personalized newsletter represents the contractual consideration that you provide for the free provision of the download. You can unsubscribe from the newsletter at any time with effect for the future. You can object to the future use of your data for advertising purposes at any time. For further information, please refer to our privacy policy.

Features

The most sensitive S and N determination thanks to two-channel system

The trace SN cube has an integrated two-channel system with two independent furnaces and combustion tubes, which makes it possible to adjust all parameters (temperature, injection volume, injection speed, etc.) for optimal sulfur and nitrogen determination with highest precision and accuracy. Samples are injected directly into the combustion tubes by the integrated 50 or 56 (temperature controllable)-position autosampler. The highly sensitive UV fluorescence and chemiluminescence detectors guarantee utmost sensitivity with detection limits of 6 ppb for sulfur and 15 ppb for nitrogen. This ultra-trace detection power gives rise to uncompromised sulfur and nitrogen analysis even in samples of lowest concentrations, and the two analyzers in one box design results in significant cost and space savings compared to the purchase of two separate analyzers.

Soot-free combustion for reliable results and low maintenance

In high-temperature combustion of oil and petrochemical samples, a well-known problem is soot formation caused by incomplete combustion. This affects data quality and also results in higher maintenance efforts. The trace SN cube successfully avoids soot formation by performing a catalyst-supported combustion, using especially developed catalysts which drastically optimize the oxidation process. With this technique, almost any sample can be ignited without soot production at injection volumes up to 80 μl, ensuring reliable results in any kind of sample and keeping maintenance as low as possible.

Highest sample flexibility

The trace SN offers outstanding sample flexibility. Apart from liquid injections with the integrated 50 (not temperature controllable) or 56 (temperature controllable)-position autosampler, the analyzer can be equipped with an optional gas, LPG, or solids module for highest versatility in your sample matrices. For the analysis of highly volatile samples we recommend the use of our temperature-controllable autosampler. Retrofitting is easily done within a few minutes.

Future-proof thanks to 10 years guarantee

Thanks to the outstanding robustness and longevity of our trace SN cube, we grant a 10-year guarantee on the high-temperature combustion furnace. With our long-term oriented dedication to technical support, we provide spare parts for a minimum of 10 years. This results in outstanding low total cost of ownership, giving you confidence in return of investment.

Product details

The trace SN cube is optimized for the analysis of:

  • SN

  • S

  • N

Sample volume

Injection volume free adjustable from 3-80 µl

Element concentration range 

Large measuring range:
S: up to 1,000 ppm, 6 ppb LOD
N: up to 1,000 ppm, 15 ppb LOD

Precision 

< 2 % rel. for 10 ppm standard (S and N)

Analysis time 

~ 4-6 min per parameter, self-optimizing according to element content and sample weight. Depending on sample type, analysis mode, and configuration.

The trace SN cube offers high-temperature combustion in the carrier gas at up to 1,150 °C, with direct injection of liquid samples into separate ports for N and S determination.

Detector 

High sensitive UV-F and CL detector, linear over entire range

Sample introduction 

Flexible sample introduction:

  • Manually with a syringe
  • Integrated 50* positions autosampler (not temperature controllable) for liquids
  • Integrated 56* positions autosampler (temperature controllable) for liquids
  • Integrated 80* or 120* positions autosampler for solids

Maximum furnace temperature 

1,200 °C 

Instrument control 

Windows® based trace SN cube operating software with LIMS integration and auto sleep and wake-up function for automated and unattended overnight operation. 21 CFR Part 11 functionality available.*

Required gases 

Synthetic air, argon, oxygen for N
Synthetic air for S

Dimensions 

47 × 55 × 57 (W × D × H) without detectors, autosampler

* optionally available

Electronic balance

Balance and elemental analyzer form a system. We offer analytical balances from the leading manufacturers with our instruments. Please contact us for further information.

SN standards dissolved in toluene

We offer a variety of SN standards dissolved toluene with concentrations from 1 µg/ml up to 1,000 µg/ml.

Downloads

Product flyer
trace SN cube Flyer
Flyer trace SN cube
download download download
Elementar Product Portfolio Brochure
Elementar Product Portfolio Brochure
download download
Cover of the organic elemental analysis application guide, incl. a man in the lab, weighing samples
Application Guide Organic Elemental Analysis
request request
Data Bulletin "Determination of low sulfur concentrations in mineral oil using the trace SN cube"
Determination of low sulfur concentrations in mineral oil using the trace SN cube
download download
Data Bulletin "Qualification of used cooking oil for recycling using the trace SN cube"
Qualification of used cooking oil for recycling using the trace SN cube
download download download
Cover of the organic elemental analysis application guide, incl. a man in the lab, weighing samples
Application Guide Organic Elemental Analysis
request request
Technical Note "Catalyst supported combustion for reliable trace element analysis"
Catalyst supported combustion for reliable trace element analysis
English
request
Can we Assist you?
Please feel free to contact us.