MAXimizing sample variety and throughput in nitrogen and protein determination
The rapid MAX N exceed PLUS is a high-performance nitrogen and protein analyzer designed for maximum laboratory efficiency and high sample throughput. Operating according to the safe, simple, and eco-friendly Dumas combustion method, it delivers fast and cost-effective nitrogen and protein determination in just five minutes per sample. Engineered for unattended 24/7 operation, its advanced crucible technology enables flexible analysis of a wide variety of sample types and weights up to 5 g / 5 ml, while optimized consumables and the use of argon as alternative carrier gas significantly reduce cost per sample. The ideal solution for laboratories with the highest demands on efficiency, flexibility, reliability, and continuous high-throughput performance.
Highlights of the rapid MAX N exceed PLUS
Maximum sample flexibility
Analysis of a wide range of sample matrices thanks to the upright crucible design and unique post-combustion technology
Best-in-class sample throughput
Fewer time-consuming calibration measurements and maintenance events mean more capacity for real sample analysis
Easy to use
Smart workflows in the intuitive software, automatic system checks, and tool-free maintenance ensure simple operation
Reliable results - anytime
Matrix-independent calibration as well as optimized sample introduction and complete combustion of the whole sample always ensure highest precision and accuracy
From Kjeldahl to Dumas:
Meeting Today’s Demands in Dairy Laboratories
Learn how dairy laboratories can achieve reliable results while increasing throughput, improving safety, and reducing operational workload in this comparative assessment of methods for protein determination.
DOWNLOAD YOUR COPY
Fill in the form to receive your download link per e-mail.
Your contractual consideration for the free provision of the download is the subscription to our personalized newsletter. By clicking on the “download now” button, you therefore declare your acceptance of the receipt of personalized newsletters by e-mail by Elementar UK Ltd. and its group companies as well as the evaluation of your user behavior in this regard and - if available - the merging of this data with your data in our customer database.
In order to receive newsletters from our group companies it is necessary to transfer your above-mentioned personal data to these companies. The data transfer is contractually required.
You are aware that the subscription to our personalized newsletter represents the contractual consideration that you provide for the free provision of the download. You can unsubscribe from the newsletter at any time with effect for the future. You can object to the future use of your data for advertising purposes at any time. For further information, please refer to our privacy policy.
Features
Maximum sample flexibility
Analyze a wide range of sample types with confidence. The upright crucible design with a capacity of up to 5 g / 5 ml allows laboratories to process a wide range of sample types - from liquids such as milk, beer, soft drinks, juice, and soy sauce to semi-liquids like yogurt and mayonnaise, and solids including meat, cereals, and flour with ease. Combined with our unique post-combustion technology, complete and unmatched combustion of even challenging sample matrices is ensured.
Best-in-class sample throughput
Quick daily maintenance routines and less time-consuming calibration measurements significantly reduce setup time – saving you time every day and increasing throughput of real samples. The 90-position autosampler can be reloaded at any time to prioritize urgent samples and offers nitrogen and protein determination in just five minutes per sample run.
Easy to use
The rapid MAX N exceed PLUS is designed for simple, reliable day-to-day operation. Improved process control and software with optimized navigation and smart workflows guide users efficiently through every step of the analysis. A status display directly on the instrument, automatic system checks, and integrated software assistants ensure quick preparation for analysis and smooth operation. Additionally, advanced EAS" REGAINER® technology, extended consumable uptime and easy, tool-free replacement of pre-filled tubes simplify routine tasks. Using the Dumas combustion method eliminates the need for hazardous chemicals, making operation safer and more convenient for laboratory staff.
Reliable results - anytime
The rapid MAX N exceed PLUS ensures outstanding precision and accuracy in nitrogen and protein determination. Thanks to the controlled oxygen jet-injection directly into the active zone and the industry-proven, catalytic post-combustion, complete combustion of the whole sample is guaranteed, even for samples hard to digest. The rapid MAX N exceed PLUS delivers matrix-independent results and offers absolute accuracy and full recovery in the low ppm range.
Future-proof with a 10-year commercial guarantee
Thanks to the outstanding robustness and longevity of the rapid MAX N exceed PLUS, we provide a 10-year commercial guarantee on the high-temperature combustion furnace and the thermal conductivity detector cell of the thermal conductivity detector (TCD). With our long-term oriented dedication to technical support, we provide spare parts for a minimum of 10 years. This results in outstanding low total cost of ownership, giving you confidence in return of investment.
Typical sample types

Product details
The rapid MAX N exceed PLUS is optimized for the analysis of:
nitrogen
protein
dietary fiber analysis*
* optionally available
Maximum sample weight
Up to 5 g or 5 ml
Element concentration range
Large concentration range:
N: up to 500 mg absolute, from 20 ppm to 100 %
Precision
helium: < 0.05 % absolute (250 mg glutamic acid)
argon: < 0.15 % absolute (250 mg glutamic acid)
Analysis time
~ 5 min, self-optimizing according to element content and sample weight
High-temperature combustion with 100 % recovery and detection of the released nitrogen. For nitrogen and protein “Dumas” analysis compliant with current ISO, EN, DIN, AOAC and other national and international standards.
Detector
Thermal conductivity detector (TCD)
Sample introduction
Sample introduction via integrated 90-position autosampler, reloadable during analysis.
Maximum furnace temperature
1,200 °C
Instrument control
Windows® based rapid MAX N exceed PLUS operating software with LIMS integration and auto sleep and wake-up function for automated and unattended overnight operation. 21 CFR Part 11 functionality available.*
Required gases
helium or argon and oxygen
Dimensions
63 x 55 x 106 cm (W x D x H)
* optionally available
Electronic balance
Balance and elemental analyzer form a system. We offer analytical balances from the leading manufacturers with our instruments. Please contact usfor further information.